Custom ECG 1 Lead Ecg Snap Cable | Raw ECG EKG Medical Cable -EC201S-005

Engineering ultra-precise, artifact-free diagnostics for intensive care monitors and ambulatory holter systems requires balancing sensitive micro-volt telemetry with rigorous physical stamina. As an agile, ISO 13485 certified raw wire innovator, Conectmed (YQF Medical Cable) custom-manufactures high-performance 1 lead ECG raw cables (Model: EC201S-005) directly on wholesale master spools. Precision-engineered for global medical instrument OEMs, this single core shielded medical wire eliminates kinetic motion artifacts while securing absolute signal fidelity under severe clinical handling.

ECG 1 Lead Ecg Snap Cable |EC201S-005
ECG 1 Lead Ecg Snap Cable |EC201S-005
EC201S-005 cutting diagram
EC201S-005 cutting diagram





technical specification

Items Specification
Products name 
ModelEC201S-005
ConductorMaterialTinned copper+Nylon
Specification28/0.08TC+500D Nylon
InsulationMaterialHDPE
Nom.Thickness0.36
Nom.O.D.ø1.40士0.05
ColourTransparent
Conducting LayerMaterialSemiconductive PVC
Nom.Thickness0.15
Specificationø1.70士0.08
ColourBlack
WrapMaterialContton Tissue
ShieldMaterialTinned copper
Specification49士3/0.10TC
Coverage>90%
JacketMaterial4394TPU
Nom.Thickness0.28
Nom.O.D.ø2.9士0.1
ColourGlossy Gray 

💡 Technical Superiority of Conectmed Low-Noise 1-Lead ECG Series 

Standard commercial wires suffer from continuous baseline drift and friction-induced microphonic artifacts whenever a patient moves. Conectmed eliminates these diagnostic errors through multi-layer hardware fortification:

* Low-Noise Semiconductive PVC Layer: Positioned perfectly between the HDPE insulation and tinned copper shield, our custom semiconductive matrix dissipates static charges caused by continuous cable friction, stabilizing your ECG/EKG waveforms and minimizing motion artifacts.
* 500D Heavy-Duty Nylon Reinforcement: The 28 AWG inner core combines ultra-fine 0.08mm tinned copper strands with 500D high-tensile Nylon (Kevlar class fillers). This prevents conductor necking or internal snapping even when nurses aggressively tug the leads in emergency rooms.
* High-Density >90% Shielding Coverage: Wrapped in a tight 49/0.10TC tinned copper braided matrix, blocking ambient high-frequency EMI from nearby surgical electrosurgical units (ESU) or hospital telemetry networks.

Contact Form